Ultrasonic Homogenizer for 100 µL to 50 mL

650L: cell disruption and direct laboratory sonication.

For homogenization, emulsification and dispersion. Stated sample range: 100 µL–50 mL; select the probe and cooling for volume, viscosity and heat sensitivity.

Price: On request

Ultrasonic Homogenizer for 100 µL to 50 mL
Category: Brand:

The Labsun 650L processes samples by direct sonication with a probe. Typical tasks include cell disruption, homogenization, emulsification and dispersion. Select the probe, vessel, pulse settings and cooling for the sample; the stated volume range does not apply to every probe.

Selection and application limits

  • Specify sample volume, viscosity, solids content and temperature limit
  • Assess probe diameter, immersion depth and vessel geometry together
  • Set power and treatment time using the actual sample and account for heating
  • Dispersion does not establish a specific nanoparticle size or chemical material identity

Technical Specifications:

  • Frequency: 20–25 kHz
  • Display: 7-inch capacitive touchscreen
  • Ultrasonic power: 650 W (1–100%)
  • Standard probe: 6 mm
  • Optional probes: 2 mm, 3 mm
  • Processing volume: 100 µL to 50 mL
  • Standard adapter: 50 mL cooling adapter
  • Optional cooling adapters: 0.2 mL, 0.5 mL, 2 mL, 5 mL, 10 mL, and 15 mL
  • Alarm temperature: 0–99.9 °C
  • Dimensions: 312 mm × 334 mm × 620 mm
  • Transducer lifting mechanism: Electric, with remote lifting function as standard
  • Interior lighting: Available
  • Cooling temperature: -4 °C to ambient temperature

 

Download Documents:

Instrument / service Description
650L – base instrument 650 W; standard Ø 6 mm probe; stated sample range 0.1–50 mL. Ø 2 mm and Ø 3 mm probes optional. Confirm suitability for the sample and vessel.
650L – annual maintenance Annual maintenance; agree scope in the quotation.

Order reference, probe configuration and scope of supply are confirmed in the quotation.

Which tasks does the 650L support?

Cell disruption, homogenization, emulsification and dispersion of suitable laboratory samples. Suitability depends on the sample, probe and treatment conditions.

Does the volume range apply to every probe?

No. The stated 100 µL–50 mL range describes the system. Probe diameter, vessel, immersion depth and cooling must suit the actual working volume.

Does dispersion guarantee a specific nanoparticle size?

No. A target particle size or material property must be checked using a separate suitable method.

What information is needed for selection?

Specify sample material, volume, viscosity, solids content, temperature limit and intended result. The quotation defines probes, cooling adapters and scope of supply.

Contact form
Enquiry / Quotation / Order
Phone:+49 (0)721 47080525
Fax:+49 (0)721 47080537
Email:info@lab-sun.de
Technical Customer Service
Phone:+49 (0) 721 47080579
Email:info@lab-sun.de