The Labsun 650L processes samples by direct sonication with a probe. Typical tasks include cell disruption, homogenization, emulsification and dispersion. Select the probe, vessel, pulse settings and cooling for the sample; the stated volume range does not apply to every probe.
Selection and application limits
- Specify sample volume, viscosity, solids content and temperature limit
- Assess probe diameter, immersion depth and vessel geometry together
- Set power and treatment time using the actual sample and account for heating
- Dispersion does not establish a specific nanoparticle size or chemical material identity
Technical Specifications:
- Frequency: 20–25 kHz
- Display: 7-inch capacitive touchscreen
- Ultrasonic power: 650 W (1–100%)
- Standard probe: 6 mm
- Optional probes: 2 mm, 3 mm
- Processing volume: 100 µL to 50 mL
- Standard adapter: 50 mL cooling adapter
- Optional cooling adapters: 0.2 mL, 0.5 mL, 2 mL, 5 mL, 10 mL, and 15 mL
- Alarm temperature: 0–99.9 °C
- Dimensions: 312 mm × 334 mm × 620 mm
- Transducer lifting mechanism: Electric, with remote lifting function as standard
- Interior lighting: Available
- Cooling temperature: -4 °C to ambient temperature
Download Documents: