Measurement task
Use low-level Pt-Co colour measurements to track yellow tint and batch-to-batch colour in nearly colourless, transparent chemicals. The Lovibond ICx is a transmission spectrophotometer for this task. Typical samples include clear solvents, liquid raw materials and process liquids. The sample specification determines which colour scale and acceptance limit apply.
Sample conditions
Use a clear, homogeneous sample without visible particles or bubbles in the light path. Review the safety data sheet and check compatibility with the sample cell and lid before measurement. Follow the specified sampling and temperature conditions so that results from different batches are comparable.
Method and colour scale
For low Pt-Co values, use the specified 100 mm optical path and a baseline and reference check appropriate to the method. Confirm the applicable instrumental procedure and reporting range, such as ISO 6271 or ASTM D5386 where specified. The ICx also reports CIE L*a*b* and Yellowness Index where these are required by the specification.
Instrument and configuration
- ICx transmission spectrophotometer (169300). For the colour of nearly colourless, transparent liquids. View the Lovibond ICx.
- W600/B/100 borosilicate glass cell (656030). 100 mm optical path for applicable low-Pt-Co measurements. Confirm suitability for the sample and the cell included in the instrument quotation. View the W600/B/100 cell.
Cells and verification materials
Use the same suitable cell and orientation for baseline and sample readings. Choose a liquid colour reference suitable for the method and target range when a liquid reference check is required. Use an instrument conformance filter for the checks defined for that filter. Confirm the reference value, certificate and handling requirements for the item ordered.
The ICx manual cites ASTM D1193 Type I or II in its ASTM D5386 section. The English application note describes ISO Grade I or II purified water. Follow the approved laboratory procedure and confirm the baseline specification for the method in use.
Application limits
Turbid, opaque or strongly coloured samples need a separate assessment of the measurement method. For corrosive media, confirm material suitability before choosing a glass cell. Colour is a separate quality parameter from purity, water content or trace contaminants. For battery formulations, refer to the transparent battery electrolyte guide.
Manufacturer documentation
- Lovibond ICx manufacturer information (English)
- ICx operating manual, issue 1 (English, PDF)
- Lovibond ICx application note (English, PDF)
Request a quotation
Tell Labsun GmbH the sample type, colour scale and expected range, applicable procedure, sample volume and temperature. Include the safety data sheet where applicable. Labsun can help specify the instrument, cell and verification materials for a quotation.
