Measurement task
Compare the colour of pharmaceutical samples with a defined reference or compare colour differences between batches. Use transmission for transparent solutions and reflectance for suitable solid surfaces. The test procedure determines the scale, sample presentation and acceptance criteria.
Sample conditions
For transparent liquids, confirm the sample volume, the cell material and the required optical path length. For tablets and capsules, check whether the available surface area and curvature suit the measurement aperture. Powders, gels and pastes need a suitable holder and a reproducible preparation procedure.
Method and colour scale
Confirm the required European Pharmacopoeia (Ph. Eur.), United States Pharmacopeia (USP) or internal test procedure before selecting the instrument. A visual comparator compares the sample with the reference set specified for the method. An instrumental configuration measures transmission or reflectance with the geometry and reporting format required for the sample.
Instrument and configuration
- Instrumental transmission. The model-specific table for PFXi-195/5 (1371955) specifies a 40 mm optical path for EP and USP scales and 50 mm for Pt-Co.
- Visual comparison. AF332 (433320) is the EP comparator configuration. AF335 (433350) is the USP configuration. Confirm the reference set and sample vessels for the chosen procedure.
- Reflectance. Compare the TR 520 (403035) and LC 100 (401100) for solid surfaces. For powders and pastes, review sample presentation and holder compatibility with the LC 100 / SV 100 kit (401120) or a complete system such as the TRA 500 (403220).
Cells and verification materials
Match the PFXi-195/5 method to the W600/OG/40 cell (606020) for the 40 mm path or the W600/OG/50 cell (606200) for the 50 mm path. Use the EP Y glass filter (131045) for its specified EP yellow-scale conformance check. Select any required liquid reference separately. The 10 mm disposable cell (132750) is listed for LC 100 / SV 100.
Application limits
Qualify the instrument, accessories and software against the specific pharmacopoeial or internal procedure. Assess reflectance results using the specified surface measurement method. Confirm the reference, software and documentation requirements for the quoted configuration.
Manufacturer documentation
- PFXi-195/5 model-specific data sheet (German, PDF)
- AF332 EP comparator data sheet (German, PDF)
- AF335 USP comparator data sheet (German, PDF)
- TR 520 reflectance spectrophotometer data sheet (German, PDF)
- LC 100 spectrocolorimeter data sheet (German, PDF)
Request a quotation
Tell Labsun GmbH the sample type and volume or usable surface area, the test procedure, target colour scale and any documentation or software requirements. Labsun can help specify the instrument, sample holder and verification materials for a quotation.