The enclosure combines airflow management, HEPA/PR filtration and monitoring for controlled weighing operations. The brochure describes AWF900 and AWF1302 configurations; the paired values below refer to those two versions respectively.
Documented configurations
| Parameter | Specification |
|---|---|
| Versions | AWF900 / AWF1302 |
| External width | 900 / 1300 mm |
| Airflow | 230 / 420 m³/h |
| Filter-module columns | 1 / 2 |
| Controls | 5-inch touchscreen; monitoring according to configuration |
Containment & selection
OEB-5 wording alone does not demonstrate a particular exposure limit. Agree substance properties, the required exposure target, handling steps, balance dimensions, airflow settings, filter-change arrangements and the relevant containment verification before use. Filter replacement requires a defined procedure; the enclosure does not by itself guarantee risk-free changeover.
Technical documents
The brochure gives reference configurations. Request the specification and containment evidence for the proposed installation.