Herolab HiCen TR Refrigerated Benchtop Laboratory Centrifuge for Larger Sample Volumes

Article No.: LS-21 10 065

The Herolab HiCen TR is a refrigerated benchtop laboratory centrifuge for larger sample volumes. With the AF 12.30 it reaches up to 15,000 rpm and 26,790 × g. The AS 4.500 swing-out rotor with the round bucket set accommodates up to 4 × 500 mL.

For applications requiring a defined chamber atmosphere, Herolab can assess a project-specific nitrogen purge configuration.

Price: On request

Herolab HiCen TR Refrigerated Benchtop Laboratory Centrifuge for Larger Sample Volumes
SKU: LS-21 10 065 Categories: , Brand:

Which configuration fits your sample volume?

For four 500 mL bottles, choose the AS 4.500 base rotor with round bucket set 20 90 502. For six 250 mL vessels, choose the AF 6.250 fixed-angle rotor. AS 4.500 trunnion ring sets accommodate 15 or 50 mL Falcon tubes. For 50 mL Falcon tubes at higher RCF, see the AF 8.50.3 Biosafe.

HiCen T with air cooling · Browse benchtop laboratory centrifuges

Technical specifications

Parameter Specification
Speed setting range 500–15,000 rpm
Cooling and temperature Setpoint range −20 °C to +40 °C. The achievable sample temperature depends on the rotor, speed, run time and starting temperature.
Dimensions (W × D × H) / net weight 721 × 670 × 490 mm / 139 kg
Power supply for this item number 230 V, 50/60 Hz, single phase

Rotors and vessel formats

Rotor and accessories Vessels Max. speed Max. RCF
AS 4.500 + 500 mL buckets
20 90 501 + 20 90 502
4 × 500 mL 4,500 rpm 3,962 × g
AF 6.250
20 22 100
6 × 250 mL 9,000 rpm 13,131 × g
AF 24.2
20 96 100
24 × 1.5 / 2.2 mL 15,000 rpm 25,155 × g
AF 12.30
20 45 100
12 × 30 mL 15,000 rpm 26,790 × g
AF 8.50.3 Biosafe
20 98 105
8 × 50 mL Falcon tubes 12,000 rpm 16,743 × g
AS 4.500 + trunnion ring set
20 90 501 + 20 90 550
4 × (7 × 50 mL Falcon) 4,500 rpm 4,093 × g
AS 4.500 + trunnion ring set
20 90 501 + 20 90 551
4 × (14 × 15 mL Falcon) 4,500 rpm 4,071 × g
AS 2.6
20 97 100
2 × 3 microplates 4,500 rpm 2,490 × g

Limits apply to the specified combination. In operation, use the lowest approved limit of the laboratory centrifuge, rotor, bucket, adapter and vessel. Confirm vessel dimensions, working volume and loading before ordering.

Two 500 mL PPCO bottles are listed for the AS 4.500 with the 500 mL round buckets: LS-25 35 51 with screw cap (Ø 98 × 114 mm) and LS-25 37 65 with a 70 mm wide mouth (Ø 98 × 112 mm). View PPCO bottles and order information

Herolab HiCen T/TR data sheet with rotors and adapters (PDF, English)

4 × 750 mL project configuration

The Labsun China catalogue lists a 4 × 750 mL configuration under 20 90 502C. We check the instrument version, rotor, buckets, bottles and operating limits for your enquiry. The 500 mL values apply to the assembly listed in the table.

Request a 4 × 750 mL configuration review

Request a HiCen configuration

HiCen TR with nitrogen purging as a project configuration

Nitrogen purging reduces the oxygen concentration in the laboratory centrifuge chamber. Herolab assesses the sample medium, rotor, vessels and operating conditions for the requested configuration. Gas supply, monitoring, accessories and scope of supply are specified in the project quotation.

For applications with explosion protection requirements, the documented suitability of the specific configuration and the required supporting documentation apply. Nitrogen purging alone does not constitute explosion protection certification.

Enquire about nitrogen purging for your project

What do I need for four 500 mL bottles?
Select AS 4.500 (20 90 501), the round bucket set (20 90 502) and suitable 500 mL bottles. Choose screw-cap or wide-mouth bottles in Ordering Information.

Which combination delivers the highest RCF?
The HiCen TR reaches up to 26,790 × g at 15,000 rpm with AF 12.30. For 500 mL bottles, see the separate AS 4.500 row for the round bucket set.

How do I choose the refrigerated version?
Choose the HiCen TR if your method requires temperature control. The setpoint range is −20 °C to +40 °C. Specify the rotor, speed, run time and starting temperature so we can assess your target sample temperature.

What should I provide for a quotation?
Include vessel type or dimensions, sample count, working volume, target RCF or speed, temperature, delivery postcode and country. For existing equipment, add the model and serial number.

Is nitrogen purging included in the standard configuration?
Herolab assesses nitrogen purging for the intended application and confirms the specific configuration in the project quotation. Provide the sample medium, vessels, required RCF or speed, temperature and available gas supply.

Contact form
Enquiry / Quotation / Order
Phone:+49 (0)721 47080525
Fax:+49 (0)721 47080537
Email:info@lab-sun.de
Technical Customer Service
Phone:+49 (0) 721 47080579
Email:info@lab-sun.de