Herolab HiCen FR Refrigerated Floor-Standing Laboratory Centrifuge

Article No.: LS-21 10 655

The Herolab HiCen FR is a refrigerated Red Line floor-standing laboratory centrifuge. With the AF 8.50.2 rotor it reaches up to 18,000 rpm and approximately 34,770 × g. Choose a configuration by vessel type and throughput or request a project review.

Price: On request

Herolab HiCen FR Refrigerated Floor-Standing Laboratory Centrifuge

Selection and installation

For 4 × 500 mL, select AS 4.500 with the round bucket set. For 6 × 250 mL, use the AF 6.250 fixed-angle rotor. For existing 15 or 50 mL Falcon tubes, use the trunnion ring sets listed below.

Temperature: −20 °C to +40 °C (setpoint range). The specified item uses a 230 V, 50/60 Hz single-phase supply. Net weight in the data sheet: 152 kg. For fixed installation or a version on rollers, tell us about the access route and installation site.

HiCen F with air cooling · Browse floor-standing laboratory centrifuges

Rotors and vessel formats

Rotor and accessories Vessels Max. speed Max. RCF
AS 4.500
20 90 501 + 20 90 502
4 × 500 mL 5,200 rpm 5,290 × g
AF 6.250
20 22 100
6 × 250 mL 10,000 rpm 16,211 × g
AF 24.2
20 96 100
24 × 1.5 / 2.2 mL 16,000 rpm 28,621 × g
AF 8.50.2
20 93 100
8 × 50 mL 18,000 rpm 34,774 × g
AF 8.50.3 Biosafe
20 98 105
8 × 50 mL Falcon tubes 13,500 rpm 21,191 × g
AS 4.500
20 90 501 + 20 90 550
4 × (7 × 50 mL Falcon) 5,200 rpm 5,466 × g
AS 4.500
20 90 501 + 20 90 551
4 × (14 × 15 mL Falcon) 5,200 rpm 5,435 × g
AS 2.6
20 97 100
2 × 3 microplates 5,000 rpm 3,075 × g

Use the approved operating limits of the ordered configuration, including the rotor, buckets, adapters and vessels.

View compatible 500 mL PPCO bottles · Herolab HiCen F/FR data sheet (PDF, English)

4 × 750 mL project configuration

The Labsun China catalogue lists a 4 × 750 mL configuration under 20 90 502C. We check the instrument version, rotor, buckets, bottles and operating limits for your enquiry. The 500 mL values apply to the assembly listed in the table.

Request a 4 × 750 mL configuration review

Request a HiCen configuration

How do I enquire about 4 × 750 mL?
Provide catalogue reference 20 90 502C and the bottles you intend to use. For an existing instrument, include the model and serial number.

Which rotor delivers the highest RCF?
The AF 8.50.2 reaches up to 34,774 × g at 18,000 rpm in the HiCen FR. See the table for the limits of each vessel configuration.

When should I choose the HiCen FR?
The FR provides controlled refrigeration. Include the rotor, speed, run time and required sample temperature in your enquiry.

What information does Labsun need?
Provide vessel type or dimensions, sample count, working volume, target RCF or speed, temperature, delivery postcode and installation requirements where relevant.

Contact form
Enquiry / Quotation / Order
Phone:+49 (0)721 47080525
Fax:+49 (0)721 47080537
Email:info@lab-sun.de
Technical Customer Service
Phone:+49 (0) 721 47080579
Email:info@lab-sun.de