Selection and installation
For 4 × 500 mL, select AS 4.500 with the round bucket set. For 6 × 250 mL, use the AF 6.250 fixed-angle rotor. For existing 15 or 50 mL Falcon tubes, use the trunnion ring sets listed below.
Temperature: −20 °C to +40 °C (setpoint range). The specified item uses a 230 V, 50/60 Hz single-phase supply. Net weight in the data sheet: 152 kg. For fixed installation or a version on rollers, tell us about the access route and installation site.
HiCen F with air cooling · Browse floor-standing laboratory centrifuges
Rotors and vessel formats
| Rotor and accessories | Vessels | Max. speed | Max. RCF |
|---|---|---|---|
| AS 4.500 20 90 501 + 20 90 502 |
4 × 500 mL | 5,200 rpm | 5,290 × g |
| AF 6.250 20 22 100 |
6 × 250 mL | 10,000 rpm | 16,211 × g |
| AF 24.2 20 96 100 |
24 × 1.5 / 2.2 mL | 16,000 rpm | 28,621 × g |
| AF 8.50.2 20 93 100 |
8 × 50 mL | 18,000 rpm | 34,774 × g |
| AF 8.50.3 Biosafe 20 98 105 |
8 × 50 mL Falcon tubes | 13,500 rpm | 21,191 × g |
| AS 4.500 20 90 501 + 20 90 550 |
4 × (7 × 50 mL Falcon) | 5,200 rpm | 5,466 × g |
| AS 4.500 20 90 501 + 20 90 551 |
4 × (14 × 15 mL Falcon) | 5,200 rpm | 5,435 × g |
| AS 2.6 20 97 100 |
2 × 3 microplates | 5,000 rpm | 3,075 × g |
Use the approved operating limits of the ordered configuration, including the rotor, buckets, adapters and vessels.
View compatible 500 mL PPCO bottles · Herolab HiCen F/FR data sheet (PDF, English)
4 × 750 mL project configuration
The Labsun China catalogue lists a 4 × 750 mL configuration under 20 90 502C. We check the instrument version, rotor, buckets, bottles and operating limits for your enquiry. The 500 mL values apply to the assembly listed in the table.






