Pyrolysis converts suitable organic samples into smaller compounds for separation and identification by GC/MS. The CDS range covers routine single-temperature analysis, programmed thermal studies and specialist high-pressure or UV-assisted experiments. The Cryo-mill is a separate sample preparation device, rather than a GC sample introduction module.
Systems and technical specifications
| Model or module | Technical overview |
|---|---|
| CDS 4000 | Single-temperature Pyroprobe; up to 1300 °C; direct mounting on the GC injection port; 9 kg. |
| CDS 6150 | DISC sample loading; programmable temperature sequences; up to 1300 °C; heated transfer line up to 350 °C; 16.3 kg. |
| CDS 6200 | DISC and Tenax-TA focusing trap as standard; reactant-gas capability; optional mass flow controller; up to 1300 °C; optional TD tube, dynamic headspace, cryotrap and external reactor modules; 17.5 kg. |
| DISC module | Vertical sample loading for compatible CDS 6150/6200 systems; programmable to 1300 °C; 1.5 kg. |
| Autosampler module | 48 positions with dynamic loading; requires DISC on a compatible 6150/6200; gas pressure greater than 15 psi; 12 kg. |
| CDS HPR | High-pressure Pyroprobe based on the 5200 platform; up to 1400 °C; maximum pressure 500 psi (3.4 MPa); catalyst reactor up to 800 °C; 25 kg. |
| Photoprobe | UV irradiation module, 260–400 nm; adjustable light intensity; for 6150/6200 with autosampler; 1 kg. |
| Cryo-mill | Cryogenic ball mill; up to 70 Hz; four 5 mL sample jars; liquid-nitrogen cooling; 25 kg. |
Temperature-program capabilities depend on the instrument and software version. The required sequence, gas handling, sample geometry and GC/MS interface are checked for the proposed configuration.